CVE-2026-7251
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password.
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (0.74%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
Eppendorf BioFlo 320 is vulnerable due to VNC server using a hard-coded password. If a remote attacker knows the network address of any BioFlo 320 model with remote access enabled, they can gain full control of the user interface by using this password. Once connected, the attacker would have full access to all control panel features for the BioFlo 320. VNC traffic is not encrypted.
- CVSS 4.0
- 9.3 CRITICALCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/M
- EPSS
- 0.74% probability · 51th percentile
- Public exploits
- None in Exploit-DB
- CISA KEV
- Not listed
- Weakness
- CWE-259
- Source
- ics-cert@hq.dhs.gov
EPSS 0.7% since 27 November 2026; no change recorded yet. Points are recorded when the score first appears and whenever it moves by a percentage point or more.