CVE-2025-54304
The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15.
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (0.50%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
An issue was discovered on Thermo Fisher Ion Torrent OneTouch 2 INS1005527 devices. When they are powered on, an X11 display server is started. The display server listens on all network interfaces and is accessible over port 6000. The X11 access control list, by default, allows connections from 127.0.0.1 and 192.168.2.15. If a device is powered on and later connected to a network with DHCP, the device may not be assigned the 192.168.2.15 IP address, leaving the display server accessible by other devices on the network. The exposed X11 display server can then be used to gain root privileges and the ability to execute code remotely by interacting with matchbox-desktop and spawning a terminal. NOTE: This vulnerability only affects products that are no longer supported by the maintainer.
- CVSS 3.1
- 9.8 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- EPSS
- 0.50% probability · 41th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-200
- Affected
- thermofisher/ion torrent onetouch 2 firmware
- Source
- cve@mitre.org
References
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.