SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityAwaiting Analysis

CVE-2025-15653

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation.

HIGH 7.0EPSS 0.17%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (0.17%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Dräger Zeus Infinity Empowered (Zeus IE) and Zeus RS C500 anesthesia workstations contain a local security vulnerability that allows unauthorized individuals with physical access to compromise software integrity via USB interface manipulation. Attackers can exploit the unprotected USB interfaces to impair therapy functions, manipulate device-processed data, or leverage the device as a pivot point for broader network-based attacks when connected to a network or Dräger Service Connect.

CVSS 4.0
7.0 HIGHCVSS:4.0/AV:P/AC:L/AT:N/PR:N/UI:N/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/M
EPSS
0.17% probability · 7th percentile
CISA KEV
Not listed
Weakness
CWE-668
Source
disclosure@vulncheck.com