CVE-2025-14096
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information.
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (0.14%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system. Other related CVE's are CVE-2025-14095 & CVE-2025-14097. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency. Required Configuration for Exposure: Attacker requires physical access to the analyzer. Temporary work Around: Only authorized people can physically access the analyzer. Permanent solution: Local Radiometer representatives will contact all affected customers to discuss a permanent solution. Exploit Status: Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.
- CVSS 3.1
- 8.4 HIGHCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- EPSS
- 0.14% probability · 3th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-250, CWE-798
- Source
- 46b595e9-1acc-41cb-9398-adaf98d37a9b
References
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.