SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityDeferred

CVE-2024-9832

An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

CRITICAL 9.3EPSS 0.18%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (0.18%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.

CVSS 3.1
9.3 CRITICALCVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
0.18% probability · 7th percentile
CISA KEV
Not listed
Weakness
CWE-307
Source
productsecurity@baxter.com

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.