CVE-2024-48966
An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to…
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (0.69%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
The software tools used by service personnel to test & calibrate the ventilator do not support user authentication. An attacker with access to the Service PC where the tools are installed could obtain diagnostic information through the test tool or manipulate the ventilator's settings and embedded software via the calibration tool, without having to authenticate to either tool. This could result in unauthorized disclosure of information and/or have unintended impacts on device settings and performance.
- CVSS 3.1
- 10.0 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
- EPSS
- 0.69% probability · 51th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-306
- Source
- productsecurity@baxter.com
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.