VulnerabilityModified
CVE-2023-30561
The data flowing between the PCU and its modules is insecure.
MEDIUM 6.1EPSS 0.19%
Does this matter?
Lower severity and a low EPSS score (0.19%). Track it; it rarely justifies an emergency change on its own.
Description
The data flowing between the PCU and its modules is insecure. A threat actor with physical access could potentially read or modify data by attaching a specially crafted device while an infusion is running.
- CVSS 3.1
- 6.1 MEDIUMCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
- EPSS
- 0.19% probability · 9th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-311
- Affected
- bd/alaris 8015 pcu firmware
- Source
- cybersecurity@bd.com
References
- https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mxMitigation, Vendor Advisory
- https://www.bd.com/en-us/about-bd/cybersecurity/bulletin/bd-alaris-system-with-guardrails-suite-mxMitigation, Vendor Advisory
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.