CVE-2023-0888
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module.
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (0.60%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
An improper neutralization of directives in dynamically evaluated code vulnerability in the WiFi Battery embedded web server in versions L90/U70 and L92/U92 can be used to gain administrative access to the WiFi communication module. An authenticated user, having access to both the medical device WiFi network (such as a biomedical engineering staff member) and the specific B.Braun Battery Pack SP with WiFi web server credentials, could get administrative (root) access on the infusion pump communication module. This could be used as a vector to start further attacks
- CVSS 3.1
- 7.2 HIGHCVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H
- EPSS
- 0.60% probability · 47th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-95, CWE-94
- Affected
- bbraun/battery-pack sp with wifi firmware
- Source
- productsecurity@bbraun.com
References
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.