CVE-2022-41627

Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data.

HIGH 7.6EPSS 0.11%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (0.11%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.

CVSS 3.1
7.6 HIGHCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:H
EPSS
0.11% probability · 1th percentile
CISA KEV
Not listed
Weakness
CWE-311, CWE-319
Affected
alivecor/kardiamobile firmware · alivecor/kardiamobile 6l firmware · alivecor/kardiamobile card firmware