SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2021-37166

A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7.

HIGH 7.5EPSS 2.06%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (2.06%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

A buffer overflow issue leading to denial of service was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. When HMI3 starts up, it binds a local service to a TCP port on all interfaces of the device, and takes extensive time for the GUI to connect to the TCP socket, allowing the connection to be hijacked by an external attacker.

CVSS 3.1
7.5 HIGHCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
EPSS
2.06% probability · 80th percentile
CISA KEV
Not listed
Weakness
CWE-120
Affected
swisslog-healthcare/hmi-3 control panel firmware
Source
cve@mitre.org

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.