A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, operated by released versions of software before Nexus Software 7.2.5.7.

CRITICAL 9.8EPSS 3.26%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (3.26%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

A buffer overflow issue was discovered in the HMI3 Control Panel contained within the Swisslog Healthcare Nexus Panel, operated by released versions of software before Nexus Software 7.2.5.7. A buffer overflow allows an attacker to overwrite an internal queue data structure and can lead to remote code execution.

CVSS 3.1
9.8 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
3.26% probability · 88th percentile
Public exploits
None in Exploit-DB
CISA KEV
Not listed
Weakness
CWE-120
Affected
swisslog-healthcare/hmi-3 control panel firmware