SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2021-37160

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7.

CRITICAL 9.8EPSS 8.23%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (8.23%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

A firmware validation issue was discovered in HMI3 Control Panel in Swisslog Healthcare Nexus Panel operated by released versions of software before Nexus Software 7.2.5.7. There is no firmware validation (e.g., cryptographic signature validation) during a File Upload for a firmware update.

CVSS 3.1
9.8 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
8.23% probability · 95th percentile
CISA KEV
Not listed
Weakness
CWE-347
Affected
swisslog-healthcare/hmi-3 control panel firmware
Source
cve@mitre.org

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.