VulnerabilityModified
CVE-2021-33885
An Insufficient Verification of Data Authenticity vulnerability in B.
CRITICAL 9.8EPSS 5.58%
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (5.58%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
An Insufficient Verification of Data Authenticity vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to send the device malicious data that will be used in place of the correct data. This results in full system command access and execution because of the lack of cryptographic signatures on critical data sets.
- CVSS 3.1
- 9.8 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
- EPSS
- 5.58% probability · 92th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-347
- Affected
- bbraun/spacecom2
- Source
- cve@mitre.org
References
- https://www.bbraunusa.com/en.htmBroken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/Exploit, Third Party Advisory
- https://www.bbraunusa.com/en.htmBroken Link
- https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/Exploit, Third Party Advisory
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.