SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2020-6966

In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme…

CRITICAL 10.0EPSS 2.22%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (2.22%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, the affected products utilize a weak encryption scheme for remote desktop control, which may allow an attacker to obtain remote code execution of devices on the network.

CVSS 3.1
10.0 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H
EPSS
2.22% probability · 82th percentile
CISA KEV
Not listed
Weakness
CWE-326
Affected
gehealthcare/apexpro telemetry server firmware · gehealthcare/carescape central station mai700 firmware · gehealthcare/carescape central station mas700 firmware · gehealthcare/clinical information center mp100d firmware · gehealthcare/clinical information center mp100r firmware · gehealthcare/carescape telemetry server mp100r firmware
Source
ics-cert@hq.dhs.gov

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.