CVE-2020-6964
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the…
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (1.36%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X and CARESCAPE Central Station (CSCS) Versions 2.X, the integrated service for keyboard switching of the affected devices could allow attackers to obtain remote keyboard input access without authentication over the network.
- CVSS 3.1
- 8.6 HIGHCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N
- EPSS
- 1.36% probability · 70th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-306
- Affected
- gehealthcare/apexpro telemetry server firmware · gehealthcare/carescape central station mai700 firmware · gehealthcare/carescape central station mas700 firmware · gehealthcare/clinical information center mp100d firmware · gehealthcare/clinical information center mp100r firmware · gehealthcare/carescape telemetry server mp100r firmware
- Source
- ics-cert@hq.dhs.gov
References
- https://www.us-cert.gov/ics/advisories/icsma-20-023-01Third Party Advisory, US Government Resource
- https://www3.gehealthcare.com/~/media/downloads/us/support/site-planning/site-readiness/gehc-gateway_project_implementation_guide_pdf.pdfProduct
- https://www.us-cert.gov/ics/advisories/icsma-20-023-01Third Party Advisory, US Government Resource
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.