CVE-2020-25187

Medtronic MyCareLink Smart 25000 is vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack.

CRITICAL 9.8EPSS 3.93%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (3.93%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Medtronic MyCareLink Smart 25000 is  vulnerable when an authenticated attacker runs a debug command, which can be sent to the patient reader and cause a heap overflow event within the MCL Smart Patient Reader software stack. The heap overflow could allow an attacker to remotely execute code on the MCL Smart Patient Reader, potentially leading to control of the device

CVSS 3.1
9.8 CRITICALCVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
3.93% probability · 90th percentile
CISA KEV
Not listed
Weakness
CWE-122, CWE-787
Affected
medtronic/mycarelink smart model 25000 firmware
Source
ics-cert@hq.dhs.gov

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.