CVE-2020-16216
In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the…
Does this matter?
Lower severity and a low EPSS score (0.70%). Track it; it rarely justifies an emergency change on its own.
Description
In IntelliVue patient monitors MX100, MX400-550, MX600, MX700, MX750, MX800, MX850, MP2-MP90, and IntelliVue X2 and X3 Versions N and prior, the product receives input or data but does not validate or incorrectly validates that the input has the properties required to process the data safely and correctly, which can induce a denial-of-service condition through a system restart.
- CVSS 3.1
- 6.5 MEDIUMCVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
- EPSS
- 0.70% probability · 51th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-20
- Affected
- philips/patient information center ix · philips/performancebridge focal point · philips/intellivue mp2-mp90 firmware · philips/intellivue mx100 firmware · philips/intellivue mx400 firmware · philips/intellivue mx850 firmware · philips/intellivue x2 firmware · philips/intellivue x3 firmware · philips/intellivue mx800 firmware · philips/intellivue mx750 firmware · philips/intellivue mx700 firmware · philips/intellivue mx600 firmware · philips/intellivue mx550 firmware
- Source
- ics-cert@hq.dhs.gov
References
- https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01Third Party Advisory, US Government Resource
- https://www.philips.com/productsecurity
- https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01Third Party Advisory, US Government Resource
- https://www.philips.com/productsecurity
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.