CVE-2020-12039
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus…
Does this matter?
Lower severity and a low EPSS score (0.33%). Track it; it rarely justifies an emergency change on its own.
Description
Baxter Sigma Spectrum Infusion Pumps Sigma Spectrum Infusion System v's6.x model 35700BAX & Baxter Spectrum Infusion System v's8.x model 35700BAX2 contain hardcoded passwords when physically entered on the keypad provide access to biomedical menus including device settings, view calibration values, network configuration of Sigma Spectrum WBM if installed.
- CVSS 3.1
- 2.4 LOWCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:N/A:N
- EPSS
- 0.33% probability · 26th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-259, CWE-798
- Affected
- baxter/sigma spectrum infusion system firmware
- Source
- ics-cert@hq.dhs.gov
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04Third Party Advisory, US Government Resource
- https://www.us-cert.gov/ics/advisories/icsma-20-170-04Third Party Advisory, US Government Resource