CVE-2020-12012
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have…
Does this matter?
Lower severity and a low EPSS score (0.34%). Track it; it rarely justifies an emergency change on its own.
Description
Baxter ExactaMix EM 2400 & EM 1200, Versions ExactaMix EM2400 Versions 1.10, 1.11, 1.13, 1.14, ExactaMix EM1200 Versions 1.1, 1.2, 1.4, 1.5, Baxter ExactaMix EM 2400 Versions 1.10, 1.11, and 1.13, and ExactaMix EM1200 Versions 1.1, 1.2, and 1.4 have hard-coded administrative account credentials for the ExactaMix application. Successful exploitation of this vulnerability may allow an attacker with physical access to gain unauthorized access to view/update system configuration or data. This could impact confidentiality and integrity of the system and risk exposure of sensitive information including PHI.
- CVSS 3.1
- 6.1 MEDIUMCVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N
- EPSS
- 0.34% probability · 28th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-259, CWE-798
- Affected
- baxter/em2400 firmware · baxter/em1200 firmware
- Source
- ics-cert@hq.dhs.gov
References
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01Third Party Advisory, US Government Resource
- https://www.us-cert.gov/ics/advisories/icsma-20-170-01Third Party Advisory, US Government Resource
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.