CVE-2019-25719
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and…
Does this matter?
High impact if exploited, but EPSS currently rates exploitation as unlikely (0.13%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.
Description
Dräger Infinity Acute Care System and Standalone Infinity M540 patient monitors running software versions VG4.1.1, VG4.0.3, and lower contain network message handling vulnerabilities that allow network-adjacent attackers to spoof or tamper with data and cause denial-of-service conditions. Attackers with access to an enabled Infinity network port or physical proximity to a wireless access point can modify device settings such as alarm states or alarm limits, and overwhelm the system with incoming data causing the device to reboot and lose network functionality.
- CVSS 4.0
- 8.8 HIGHCVSS:4.0/AV:N/AC:L/AT:N/PR:N/UI:N/VC:L/VI:L/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/M
- EPSS
- 0.13% probability · 3th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-924
- Source
- disclosure@vulncheck.com
References
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.