SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2018-8868

An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality.

MEDIUM 6.2EPSS 0.33%

Does this matter?

Lower severity and a low EPSS score (0.33%). Track it; it rarely justifies an emergency change on its own.

Description

Medtronic 24950 MyCareLink Monitor and 24952 MyCareLink Monitor contains debug code meant to test the functionality of the monitor's communication interfaces, including the interface between the monitor and implantable cardiac device. An attacker with physical access to the device can exploit other vulnerabilities to access this debug functionality. This debug functionality provides the ability to read and write arbitrary memory values to implantable cardiac devices via inductive or short range wireless protocols. An attacker with close physical proximity to a target implantable cardiac device can use this debug functionality.

CVSS 3.1
6.2 MEDIUMCVSS:3.1/AV:P/AC:H/PR:L/UI:N/S:C/C:H/I:L/A:L
EPSS
0.33% probability · 26th percentile
CISA KEV
Not listed
Weakness
CWE-749
Affected
medtronic/24950 mycarelink monitor firmware · medtronic/24952 mycarelink monitor firmware
Source
ics-cert@hq.dhs.gov

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.