CVE-2018-14781
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack.
Does this matter?
Lower severity and a low EPSS score (0.71%). Track it; it rarely justifies an emergency change on its own.
Description
Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.
- CVSS 3.1
- 5.3 MEDIUMCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
- EPSS
- 0.71% probability · 52th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-294, CWE-287
- Affected
- medtronicdiabetes/508 minimed insulin pump firmware · medtronicdiabetes/522 paradigm real-time firmware · medtronicdiabetes/722 paradigm real-time firmware · medtronicdiabetes/523 paradigm revel firmware · medtronicdiabetes/723 paradigm revel firmware · medtronicdiabetes/523k paradigm revel firmware · medtronicdiabetes/723k paradigm revel firmware · medtronicdiabetes/551 minimed 530g firmware · medtronicdiabetes/751 minimed 530g firmware
- Source
- ics-cert@hq.dhs.gov
References
- http://www.securityfocus.com/bid/105044Third Party Advisory, VDB Entry
- https://global.medtronic.com/xg-en/product-security/security-bulletins/minimed.html
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02Third Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/105044Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02Third Party Advisory, US Government Resource
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.