SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2018-14781

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack.

MEDIUM 5.3EPSS 0.71%

Does this matter?

Lower severity and a low EPSS score (0.71%). Track it; it rarely justifies an emergency change on its own.

Description

Medtronic MiniMed MMT devices when paired with a remote controller and having the “easy bolus” and “remote bolus” options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery.

CVSS 3.1
5.3 MEDIUMCVSS:3.1/AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:H/A:N
EPSS
0.71% probability · 52th percentile
CISA KEV
Not listed
Weakness
CWE-294, CWE-287
Affected
medtronicdiabetes/508 minimed insulin pump firmware · medtronicdiabetes/522 paradigm real-time firmware · medtronicdiabetes/722 paradigm real-time firmware · medtronicdiabetes/523 paradigm revel firmware · medtronicdiabetes/723 paradigm revel firmware · medtronicdiabetes/523k paradigm revel firmware · medtronicdiabetes/723k paradigm revel firmware · medtronicdiabetes/551 minimed 530g firmware · medtronicdiabetes/751 minimed 530g firmware
Source
ics-cert@hq.dhs.gov

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.