SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2017-5149

This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

HIGH 8.9EPSS 1.24%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (1.24%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

An issue was discovered in St. Jude Medical Merlin@home, versions prior to Version 8.2.2 (RF models: EX1150; Inductive models: EX1100; and Inductive models: EX1100 with MerlinOnDemand capability). The identities of the endpoints for the communication channel between the transmitter and St. Jude Medical's web site, Merlin.net, are not verified. This may allow a man-in-the-middle attacker to access or influence communications between the identified endpoints.

CVSS 3.1
8.9 HIGHCVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:L/I:H/A:H
EPSS
1.24% probability · 67th percentile
CISA KEV
Not listed
Weakness
CWE-476
Affected
abbott/merlin\@home firmware
Source
ics-cert@hq.dhs.gov

References

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.