CVE-2017-12726

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.

HIGH 7.3EPSS 1.08%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (1.08%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

A Use of Hard-coded Password issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. Telnet on the pump uses hardcoded credentials, which can be used if the pump is configured to allow external communications. Smiths Medical assesses that it is not possible to upload files via Telnet and the impact of this vulnerability is limited to the communications module.

CVSS 3.0
7.3 HIGHCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L
EPSS
1.08% probability · 63th percentile
CISA KEV
Not listed
Weakness
CWE-798
Affected
smiths-medical/medfusion 4000 wireless syringe infusion pump