CVE-2017-12725
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.
Does this matter?
Lower severity and a low EPSS score (0.83%). Track it; it rarely justifies an emergency change on its own.
Description
A Use of Hard-coded Credentials issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump with default network configuration uses hard-coded credentials to automatically establish a wireless network connection. The pump will establish a wireless network connection even if the pump is Ethernet connected and active; however, if the wireless association is established and the Ethernet cable is attached, the pump does not attach the network stack to the wireless network. In this scenario, all network traffic is instead directed over the wired Ethernet connection.
- CVSS 3.0
- 5.6 MEDIUMCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:L/I:L/A:L
- EPSS
- 0.83% probability · 56th percentile
- CISA KEV
- Not listed
- Weakness
- CWE-798
- Affected
- smiths-medical/medfusion 4000 wireless syringe infusion pump
- Source
- ics-cert@hq.dhs.gov
References
- http://www.securityfocus.com/bid/100665Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02AThird Party Advisory, US Government Resource
- http://www.securityfocus.com/bid/100665Third Party Advisory, VDB Entry
- https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02AThird Party Advisory, US Government Resource
Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.