An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.

MEDIUM 5.3EPSS 2.42%

Does this matter?

Lower severity and a low EPSS score (2.42%). Track it; it rarely justifies an emergency change on its own.

Description

An Out-of-bounds Read issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. A third-party component used in the pump reads memory out of bounds, causing the communications module to crash. Smiths Medical assesses that the crash of the communications module would not impact the operation of the therapeutic module.

CVSS 3.0
5.3 MEDIUMCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:L
EPSS
2.42% probability · 83th percentile
CISA KEV
Not listed
Weakness
CWE-125
Affected
smiths-medical/medfusion 4000 wireless syringe infusion pump