SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2017-12721

An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6.

MEDIUM 5.9EPSS 0.67%

Does this matter?

Lower severity and a low EPSS score (0.67%). Track it; it rarely justifies an emergency change on its own.

Description

An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack.

CVSS 3.0
5.9 MEDIUMCVSS:3.0/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.67% probability · 52th percentile
Public exploits
None in Exploit-DB
CISA KEV
Not listed
Weakness
CWE-295
Affected
smiths-medical/medfusion 4000 wireless syringe infusion pump
Source
ics-cert@hq.dhs.gov
EPSS trend
0%3%5%27 October 2026: 0.67%27 October 202627 October 2026

EPSS 0.7% since 27 October 2026; no change recorded yet. Points are recorded when the score first appears and whenever it moves by a percentage point or more.

References

Source: NVD record, EPSS from FIRST.org, KEV from CISA, exploits from Exploit-DB. Refreshed daily. Download this record as JSON.