SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2017-12716

Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

MEDIUM 6.5EPSS 0.24%

Does this matter?

Lower severity and a low EPSS score (0.24%). Track it; it rarely justifies an emergency change on its own.

Description

Abbott Laboratories Accent and Anthem pacemakers manufactured prior to Aug 28, 2017 transmit unencrypted patient information via RF communications to programmers and home monitoring units. Additionally, the Accent and Anthem pacemakers store the optional patient information without encryption. CVSS v3 base score: 3.1, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:L/I:N/A:N. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CVSS 3.0
6.5 MEDIUMCVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.24% probability · 16th percentile
CISA KEV
Not listed
Weakness
CWE-311, CWE-319
Affected
abbott/accent firmware · abbott/anthem firmware · abbott/accent mri firmware · abbott/accent st firmware
Source
ics-cert@hq.dhs.gov

References

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.