SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2017-12714

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker…

MEDIUM 6.5EPSS 0.63%

Does this matter?

Lower severity and a low EPSS score (0.63%). Track it; it rarely justifies an emergency change on its own.

Description

Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted "RF wake-up" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities.

CVSS 3.0
6.5 MEDIUMCVSS:3.0/AV:A/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H
EPSS
0.63% probability · 50th percentile
Public exploits
None in Exploit-DB
CISA KEV
Not listed
Weakness
CWE-920
Affected
abbott/accent firmware · abbott/anthem firmware · abbott/accent mri firmware · abbott/accent st firmware · abbott/assurity firmware · abbott/allure firmware · abbott/assurity mri firmware
Source
ics-cert@hq.dhs.gov
EPSS trend
0%3%5%27 October 2026: 0.63%27 October 202627 October 2026

EPSS 0.6% since 27 October 2026; no change recorded yet. Points are recorded when the score first appears and whenever it moves by a percentage point or more.

References