SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2015-3956

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes…

CRITICAL 9.8EPSS 0.95%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (0.95%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior accept drug libraries, firmware updates, pump commands, and unauthorized configuration changes from unauthenticated devices on the host network. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 3.0
9.8 CRITICALCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
0.95% probability · 59th percentile
CISA KEV
Not listed
Weakness
CWE-345
Affected
pifzer/plum a\+ infusion system firmware · pifzer/plum a\+3 infusion system firmware · pifzer/symbiq infusion system firmware
Source
ics-cert@hq.dhs.gov

References

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.