An unauthorized user could issue commands to the pump.

CRITICAL 9.8EPSS 2.04%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (2.04%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 3.0
9.8 CRITICALCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
2.04% probability · 80th percentile
CISA KEV
Not listed
Weakness
CWE-285
Affected
pifzer/plum a\+ infusion system firmware · pifzer/plum a\+3 infusion system firmware · pifzer/symbiq infusion system firmware