SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2015-3952

Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

HIGH 7.5EPSS 1.01%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (1.01%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue.

CVSS 3.0
7.5 HIGHCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
1.01% probability · 61th percentile
CISA KEV
Not listed
Weakness
CWE-312, CWE-200
Affected
pifzer/plum a\+ infusion system firmware · pifzer/plum a\+3 infusion system firmware · pifzer/symbiq infusion system firmware
Source
ics-cert@hq.dhs.gov

References