CVE-2015-1012

Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities.

HIGH 7.5EPSS 0.77%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (0.77%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access.

CVSS 3.0
7.5 HIGHCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N
EPSS
0.77% probability · 54th percentile
CISA KEV
Not listed
Weakness
CWE-312, CWE-200
Affected
pfizer/lifecare pca infusion system firmware
Source
ics-cert@hq.dhs.gov