SOC status:Duty analyst on shift

UK Cyber Defence
VulnerabilityModified

CVE-2014-5434

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol.

CRITICAL 9.8EPSS 1.56%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (1.56%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16 has a default account with hard-coded credentials used with the FTP protocol. Baxter asserts no files can be transferred to or from the WBM using this account. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

CVSS 3.0
9.8 CRITICALCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
1.56% probability · 74th percentile
CISA KEV
Not listed
Weakness
CWE-259, CWE-798
Affected
baxter/sigma spectrum infusion system firmware
Source
ics-cert@hq.dhs.gov

References

Source: NVD record, EPSS from FIRST.org, KEV from CISA. Refreshed daily.