An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16,…

CRITICAL 9.8EPSS 2.06%

Does this matter?

High impact if exploited, but EPSS currently rates exploitation as unlikely (2.06%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit.

Description

An unauthenticated remote attacker may be able to execute commands to view wireless account credentials that are stored in cleartext on Baxter SIGMA Spectrum Infusion System version 6.05 (model 35700BAX) with wireless battery module (WBM) version 16, which may allow an attacker to gain access the host network. Baxter has released a new version of the SIGMA Spectrum Infusion System, Version 8, which incorporates hardware and software changes.

CVSS 3.0
9.8 CRITICALCVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H
EPSS
2.06% probability · 80th percentile
CISA KEV
Not listed
Weakness
CWE-312, CWE-255
Affected
baxter/sigma spectrum infusion system firmware