{"id":"CVE-2024-48974","url":"https://www.cyber-defence.io/tools/cve/CVE-2024-48974","generatedAt":"2026-10-28T00:51:48.901Z","title":"This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file.","description":"The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure.","published":"2024-11-14T22:15:19.000Z","lastModified":"2026-06-17T07:59:09.000Z","status":"Deferred","sourceIdentifier":"productsecurity@baxter.com","cvss":{"version":"3.1","score":9.3,"severity":"CRITICAL","vector":"CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H"},"cwe":["CWE-494"],"affected":[],"epss":{"score":0.00154,"percentile":0.04471,"date":"2026-10-26","history":[]},"kev":{"listed":false},"exploits":{"count":0,"verified":false,"firstPublished":null,"source":"Exploit-DB (https://gitlab.com/exploit-database/exploitdb)","entries":[]},"verdict":{"level":"high","text":"High impact if exploited, but EPSS currently rates exploitation as unlikely (0.15%). Schedule it in the normal patch cycle and watch for a rise in EPSS or a public exploit."},"changes":[],"references":[{"url":"https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01","source":"productsecurity@baxter.com"}],"sources":{"nvd":"https://nvd.nist.gov/vuln/detail/CVE-2024-48974","epss":"https://www.first.org/epss/","kev":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","exploitdb":"https://www.exploit-db.com/"},"licence":"CC BY 4.0 — link back to the CVE Explorer if you publish the results; upstream data remains subject to its own terms."}