{"id":"CVE-2018-14801","url":"https://www.cyber-defence.io/tools/cve/CVE-2018-14801","generatedAt":"2026-10-28T21:18:56.573Z","title":"In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser password and physical access can enter the superuser password that can be used to access and modify all settings on the…","description":"In Philips PageWriter TC10, TC20, TC30, TC50, TC70 Cardiographs, all versions prior to May 2018, an attacker with both the superuser password and physical access can enter the superuser password that can be used to access and modify all settings on the device, as well as allow the user to reset existing passwords.","published":"2018-08-22T18:29:01.000Z","lastModified":"2026-06-17T01:41:40.000Z","status":"Modified","sourceIdentifier":"ics-cert@hq.dhs.gov","cvss":{"version":"3.0","score":6.2,"severity":"MEDIUM","vector":"CVSS:3.0/AV:P/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H"},"cwe":["CWE-798"],"affected":["philips/pagewriter tc70 firmware","philips/pagewriter tc50 firmware","philips/pagewriter tc30 firmware","philips/pagewriter tc20 firmware","philips/pagewriter tc10 firmware"],"epss":{"score":0.00408,"percentile":0.349,"date":"2026-10-27","history":[]},"kev":{"listed":false},"exploits":{"count":0,"verified":false,"firstPublished":null,"source":"Exploit-DB (https://gitlab.com/exploit-database/exploitdb)","entries":[]},"verdict":{"level":"low","text":"Lower severity and a low EPSS score (0.41%). Track it; it rarely justifies an emergency change on its own."},"changes":[],"references":[{"url":"http://www.securityfocus.com/bid/105103","tags":["Third Party Advisory","VDB Entry"],"source":"ics-cert@hq.dhs.gov"},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-228-01","tags":["Third Party Advisory","US Government Resource","VDB Entry"],"source":"ics-cert@hq.dhs.gov"},{"url":"https://www.usa.philips.com/healthcare/about/customer-support/product-security","tags":["Vendor Advisory"],"source":"ics-cert@hq.dhs.gov"},{"url":"http://www.securityfocus.com/bid/105103","tags":["Third Party Advisory","VDB Entry"],"source":"af854a3a-2127-422b-91ae-364da2661108"},{"url":"https://ics-cert.us-cert.gov/advisories/ICSMA-18-228-01","tags":["Third Party Advisory","US Government Resource","VDB Entry"],"source":"af854a3a-2127-422b-91ae-364da2661108"},{"url":"https://www.usa.philips.com/healthcare/about/customer-support/product-security","tags":["Vendor Advisory"],"source":"af854a3a-2127-422b-91ae-364da2661108"}],"sources":{"nvd":"https://nvd.nist.gov/vuln/detail/CVE-2018-14801","epss":"https://www.first.org/epss/","kev":"https://www.cisa.gov/known-exploited-vulnerabilities-catalog","exploitdb":"https://www.exploit-db.com/"},"licence":"CC BY 4.0 — link back to the CVE Explorer if you publish the results; upstream data remains subject to its own terms."}